Intake Health

InFlow pin test

Run the test and compare the front and top lights with the two units in the video.

How to run the test

  1. Hold the unit as shown. Wrap one hand around the front and side, keeping the top edge visible.
  2. Press the test pin with your thumb.
  3. Watch the front light and then the top edge. Compare what you see with the examples below.

What to look for

Working example showing an amber glow at the top edge after the front light turns off

Working unit

The front light comes on brightly and then goes out. The top edge then shows an amber glow.

Faulty example showing a small light that remains on at the front

Faulty unit

The front stays partially lit and the top gives a brief white flash. If your unit matches this example, set it aside for replacement.

Read the transcript

If you wrap your hand around the front like that, put your hand on the side and press the pin, it’ll light up. Run a test and you should see a result like this. This is good.

On a unit that’s not working, do the same thing. Wrap your hand around it, press it, put your thumb on it. It lights up. You can see this stays half lit. It didn’t work real well, and the top just flashed white. This is wrong. I’ll show you again the top.

Check the top out. That means it’s broken. Keep that one and we will replace it for you.